Quantity of liquidQuantity of liquid :13, among which 1 to 9 areused to contain commondehydration solvents and 14 to16 to contain cleaning agents
Quantity of paraffin cells:3, numbered 10, 12 and 13
Quantity of working cells:1
Quantity of waste liquid cells:1
Volume of working chamber:9L
Volume of solvent container:6L
Temperature of working chamber:
≤45℃ (adjustable) when the medium is solvent
58~70℃ (adjustable) when the medium is paraffin
Working pressure:
Positive pressure <30 kpa
Negative pressure 0 to -70 kpa
Temperature of paraffin cell:
58~70℃ (adjustable)
Melting time of paraffin:≤3h
Rated power:1200VA
Setting time of dipping bath:0-300min (adjustable)
Time for charing/discharging liquid:0-5min
Stirring time:2min (adjustable)
Stirring interval:8min (preset)
Mode of operation:Manual and automatic
Dimensions:650 × 650 × 1030 mm (S*W*H)
Weight:145kg
Unique Gas Pressure Conversion Device Design
The gas pressure conversion originally completed by the electromagnetic valve is completed presently by the mechanical structure device, avoiding the high failure rate caused by the electromagnetic valve.
One Instrument Dual Purpose
One-button-free switch between conventional dehydration and rapid dehydration. The rapid dehydration program is started during the day to dehydrate small samples rapidly in 3 hours. The conventional dehydration program is started at night.
Environmental Protection
The tissue processing system is closed and provided with independent liquid and gas passages. The gas in the gas passage can be liquefied and discharged into the waste liquid container through the specially designed condenser pipe so that no polluted gas leaks out. It meets the requirements of environmental protection.
Multiple Safety Protection Functions
Two optical sensors are provided to sense the lowest and highest liquid levels. Prompt and service stop protection functions; working chamber over-temperature detection and power-off protection functions; automatic alarm function of unlocked working chamber.
Reagent Management System
The service cycle of reagents and paraffin is set according to the number of embedding boxes processed, the number of days of reagent use or the number of dehydration times. The system counts automatically and gives an automatic prompt to replace the reagents and paraffin after the threshold is reached.
Efficient and Stable
300 to 350 standard embedding boxes arranged orderly can be processed each time. samples are processed with reagent movement but without sample movement, which can avoid the danger of sample damage caused by power failure of the dehydrator, cylinder or other mechanical failures.
Preset Tissue Processing Program
There are 10 sets of tissue processing programs for users to set and choose from.
Appointment Function
With the function of timed automatic operation, the instrument allows users to set the completion time of tissue dehydration on any day by week.
Power-off Protection Function
With the power-off protection function, the instrument can work in sequence from the breakpoint automatically after re-connection of power.
Auto Remote Telephone Alarm Function
Timely equipment troubleshooting (optional).
Job Done Prompt
Voice and screen prompts at the end of work.
Stirring Function
Several stirring modes such as normal-pressure, pressurized, vacuum and magnetic stirring are available. Users can set the stirring time and interval and whether to enable the stirring function.
Paraffin or Reagent Auto Rotation Function
The instrument will remind users of replacement automatically after the threshold value is reached according to the effective threshold value of use of paraffin or reagents. The system will rotate paraffin or reagents automatically in the ascending order of cleanliness. Waste paraffin or liquid will be discharged automatically to the discharge box. You only need to add new paraffin or reagents to the empty position.